ESCRS - PP21.09 - Electrostatic Failure At The Posterior Capsule - Intraocular Lens Interface: An Underrecognized Mechanism In Posterior Capsular Opacification

Electrostatic Failure At The Posterior Capsule - Intraocular Lens Interface: An Underrecognized Mechanism In Posterior Capsular Opacification

Published 2026 - 44th Congress of the ESCRS

Reference: PP21.09 | Type: Poster | DOI: 10.82333/bcnq-g961

Authors: Mr Panos Gartaganis* 1 , Prof. Sotirios Gartaganis 2 , Prof. Petros Koutsoukos 3

1Department of Ophthalmology,Mediterraneo Hospital,Athens,Greece;Department of Chemical Engineering, Laboratory of Inorganic and Analytical Chemistry,University of Patras and FORTH-ICEHT,Patras,Greece, 2School of Medicine, Department of Ophthalmology,University of Patras,Patras,Greece, 3Department of Chemical Engineering, Laboratory of Inorganic and Analytical Chemistry,University of Patras and FORTH-ICEHT,Patras,Greece

Purpose

The purpose of this study is to evaluate how electrostatic surface charge interactions between the posterior lens capsule and intraocular lens (IOL) biomaterials contribute to the development of Posterior Capsular Opacification (PCO).

Setting

This study was conducted at the Laboratory of Inorganic and Analytical Chemistry, Department of Chemical Engineering, University of Patras, and at FORTH-ICEHT (Foundation for Research and Technology–Hellas, Institute of Chemical Engineering Sciences), Patras, Greece.

Methods

A mechanistic model of capsule-IOL interaction was developed based on the biochemical composition of the posterior capsule and the physicochemical surface characteristics of contemporary acrylic IOLs. The posterior capsule contains type IV collagen, laminin, and proteoglycans which give it a net negative charge in the aqueous environment. Similarly, most acrylic IOLs have a negative surface charge; however, the degree of the surface charge depends upon the composition of the polymer and its water content. A basic principle of electrostatics was applied in combination with the previously described mechanisms of capsular bend formation and lens epithelial cell (LEC) migration.

Results

Like-charge electrostatic repulsion between the posterior capsule and negatively charged IOL surfaces may reduce capsule - IOL adhesion, promoting formation of a micro-interface space at the optic - capsule junction. This space facilitates residual lens epithelial cell migration, proliferation, and extracellular matrix deposition, which are central mechanisms in PCO pathogenesis. Hydrophilic IOLs, characterized by higher water content, may exhibit stronger electrostatic effects and weaker capsular adhesion. Hydrophobic IOLs partially compensate through sharper posterior optic edges and enhanced capsular bend formation, creating a mechanical barrier to limit cellular migration and proliferation.

Conclusions

Electrostatic repulsion is an unappreciated biological factor that contributes to the development of PCO. Thus, optimizing the surface charge characteristics of IOLs may provide a novel approach to improving capsule-IOL adhesion and subsequently reducing the incidence of PCO over time after cataract surgery.