Using Corneal Wavefront For Astigmatism Prediction In Cataract Surgery
Published 2026 - 44th Congress of the ESCRS
Reference: PP17.08 | Type: Free paper | DOI: 10.82333/tej9-9c71
Authors: Dr Bjørn Gjerdrum* 1 , Dr Kjell Gunnar Gundersen 2
1ifocus øyeklinikk,Ifocus øyeklinikk,Haugesund,Norway;Ifocus øyeklinikk,Ifocus øyeklinikk,Haugesund,Norway, 2Ifocus øyeklinikk,Ifocus øyeklinikk,Haugesund,Norway
Purpose
To evaluate astigmatism prediction error using preoperative total corneal wavefront measurements.
Setting
The study included results from two private eye clinics situated in Haugesund and Stavanger in Norway
Methods
This was a retrospective study of successful cataract or refractive lens exchange surgery with toric and non-toric IOLS. Preoperative biometry was performed with an OCT based biometer (Anterion, Heidelberg Engineering. The astigmatism prediction error (APE) was calculated as the magnitude of the vector differences between the achieved and the predicted refractive astigmatism using thre methods: Built-in Barrett toric calclulator (measured posterior astigmatism) (APE-Barrett), total cornea power (APE-TCP) and total wavefront astigmatism (APE-WF). Wavefront astigmatism was converted from Zernike coefficients (microns) of oblique and with-the-rule/against-the-rule astigmatism to diopters of cylinder and axis)
Results
Conclusions
Astigmatism prediction using OCT based total corneal wavefront could not improve astigmatism prediction in this study. The lowest astigmatism prediction error was found with the Barrett toric calculator for measured posterior astigmatism