One Technique, Different Iol Challenges: Adaptive Flanged Fixation In Zonular Insufficiency
Published 2026 - 44th Congress of the ESCRS
Reference: PP09.08 | Type: Free paper | DOI: 10.82333/2nt4-zp65
Authors: Dr Fabio Vicente Zavarse Fadul* 1 , Dr Naon Kim Yeon 2 , Dr Alicia Garde González 3 , Dr David Antolín García 4
1Head of the Ophthalmology department,Hospital Universitario de Torrejón,Madrid,Spain;Ophthalmology department,Clínica Central Ocular,Madrid ,Spain, 2Ophthalmology ,Hospital Universitario de Torrejón,Madrid ,Spain, 3Ophthalmology ,Clínica Central Ocular,Madrid ,Spain, 4Head of the Ophthalmology department,Clínica Central Ocular,Madrid ,Spain
Purpose
To describe technical adaptations of a modified Canabrava fixation technique for stabilizing different intraocular lens (IOL) designs in the setting of zonular insufficiency.
Setting
Private and Public ophthalmology practice.
Hospital universitario de Torrejón (public)
Clínica Central Ocular (private)
Methods
Three eyes with lens instability were managed using tailored modifications of a flange-based scleral fixation approach
In a 10-year subluxated capsular bag-PMMA IOL complex, fixation was achieved by positioning the flange loop closer to the haptic-optic junction due to the absence of a conventional anchoring zone
In a previously vitrectomized eye with zonular weakness following retinal detachment surgery, an Akreos Adapt IOL that dropped into the vitreous cavity was repositioned and stabilized
Results
Stable centration was achieved in all cases.
In the PMMA case, the patient had preoperative vision of 0.4 only in upright gaze, with no functional vision in other positions. Postoperatively, uncorrected visual acuity improved to 0.7 at one week and 0.8 with correction.
The Akreos IOL remained well centered despite limited visual prognosis due to prior retinal pathology.
The toric FineVision IOL achieved 0.94 distance and 1.0 near visual acuity with maintained axis stability.
No early tilt, decentration, or fixation-related complications were observed.
Conclusions
Targeted adaptation of flange positioning and fixation geometry enabled effective stabilization across rigid PMMA, eyelet-based, and toric premium IOL platforms. Compensation of residual zonular forces through tailored suture tension may expand surgical options for managing complex lens instability in diverse clinical scenarios. The versatility of this approach, including its use in acute intraoperative complications, may support favorable outcomes even in demanding premium toric IOLs.