Clinical Validation Of A Novel Digital Intraoperative Guidance System For Accurate Toric Iol Alignment
Published 2026 - 44th Congress of the ESCRS
Reference: PP09.05 | Type: Free paper | DOI: 10.82333/snmf-7g31
Authors: Dr Clayton Blehm* 1 , Dagny Zhu 2
1North Georgia Eye Associates,Gainesville,United States, 2Modern Laser & Cataract Eye Surgery,Zhu Vision,Orange,United States
Purpose
To evaluate the intraoperative alignment accuracy of a novel digital guidance system for toric intraocular lens (IOL) implantation and to assess1-month postoperative refractive outcomes.
Setting
Single-surgeon, single-center private practice cataract surgery facility
Methods
Prospective, observational study of 35 consecutive eyes undergoing toric IOL implantation. The digital guidance system (Cassini Technologies, B.V.) calculated and displayed intraoperative alignment axes. Three independent masked reviewers annotated paired pre- and intraoperative images using six anatomical landmarks to determine a reference axis. Eyes with ≥3° inter-reviewer disagreement were excluded. The primary endpoint was absolute difference between displayed and reference axes. Secondary endpoints included time to guidance initiation and 1-month refractive outcomes.
Results
Nineteen eyes met agreement criteria for primary analysis. Mean absolute axis difference was 1.51° ± 1.84° (median 0.59°). Mean time to guidance initiation was 17.6 seconds (n=35). At 1 month follow-up, 13 eyes were available for refractive analysis; mean residual refractive cylinder was 0.31 ± 0.34 D. Mean uncorrected distance visual acuity was 0.15 ± 0.12 logMAR (logarithm of the minimum angle of resolution) and corrected distance visual acuity was 0.05 ± 0.09 logMAR.
Conclusions
The digital guidance system demonstrated strong agreement with an anatomical reference standard for toric IOL axis determination. Postoperative outcomes showed low residual refractive astigmatism and favorable visual acuity, supporting clinically meaningful intraoperative alignment precision.