ESCRS - POCAT599 - Refractive Outcomes With Argos Biometer And Image Guidance In Post-Lasik Cataract Surgery

Refractive Outcomes With Argos Biometer And Image Guidance In Post-Lasik Cataract Surgery

Published 2026 - 44th Congress of the ESCRS

Reference: POCAT599 | Type: Free paper | DOI: 10.82333/0rxd-pj64

Authors: Mr Tzu Te Hu* 1

1Nobel Eye Institude,Taipei,Taiwan

Purpose

To evaluate the accuracy of refractive prediction using an Argos biometer with Alcon image guidance in post-myopic LASIK patients undergoing cataract surgery.

Setting

Taiwan Nobel Eye Institude

Methods

This prospective study included 60 eyes implanted with Clareon™ Aspheric or Toric IOLs. The Barrett True-K Toric calculator was used. Primary endpoints were mean absolute error (MAE) and percentage of eyes within ±0.50 D at 1 month. Secondary endpoints included residual astigmatism within ±0.25, ±0.50, ±0.75, and ±1.00 D, refractive prediction error (RPE) within ±0.25, ±0.50, ±0.75, and ±1.00 D, and visual acuity (BCDVA and DCIVA). 

Results

Interim analysis of 20 eyes showed MAE of 0.56 ± 0.58 D. 70% of eyes achieved RPE within ±0.50 D. Residual astigmatism within ±0.25 D, ±0.50 D, ±0.75 D, and ±1.00 D was 40%, 65%, 75%, and 90%, respectively. RPE within ±0.25 D, ±0.50 D, ±0.75 D, and ±1.00 D was 45%, 70%, 80%, and 80%. BCDVA and DCIVA were 0.03 ± 0.065 logMAR and 0.21 ± 0.10 logMAR. 

Conclusions

The Argos biometer demonstrates promising accuracy for IOL power calculation in post-myopic LASIK eyes. Further data collection is ongoing.