Purpose

To compare the optical function of Synergy and Odyssey intraocular lens by mobile model eye

Setting

Yeoudi St. Mary's hospital, Seoul, Korea

Methods

The mobile model eye consists of an artificial cornea, an IOL, a wet cell, an adjustable lens tube, an objective lens, and a digital single-lens reflex camera. We collected photographs of distant buildings (day and night) and streets at night, and videos of United States Air Force resolution target from 6 m to 25 cm and analyzed them quantitatively. We compared the results of Synergy (Johnson&Johnson) and Odyssey(Johnson&Johnson)IOL.

Results

The letters on the signboard of distant buildings looked clearer in Odyssey than in Synergy. For Synergy, the contrast modulation was 0.104 at 6 m. It then decreased and increased again. At 47 cm, it was 0.058 and then decreased again. For Odyssey, the contrast modulation was 0.156 at 6 m. It then decreased and increased again. At 47 cm, it was 0.064 and then decreased again. Halo around light at night was comparable in Synergy and Odyssey.

Conclusions

Contrast modulation of Odyssey is higher than that of Synergy at far distance. Halos around light at night were comparable in Synergy and Odyssey.