ESCRS - POCAT460 - Optical And Clinical Outcomes Of A New Diffractive Trifocal Intraocular Lens

Optical And Clinical Outcomes Of A New Diffractive Trifocal Intraocular Lens

Published 2026 - 44th Congress of the ESCRS

Reference: POCAT460 | Type: Free paper | DOI: 10.82333/1vvy-ky25

Authors: Dr Belen Alfonso-Bartolozzi* 1 , Dr Luis Fernandez-Vega-Cueto 1 , Dr Carlos Lisa Fernández 1 , Dr Ignacio Rodriguez-Uña 1 , Prof. David Madrid Costa 1 , Dr José F. Alfonso Sanchez 1

1Fernández-Vega Ophthalmological Institute,IOFV,Oviedo,Spain

Purpose

To evaluate the clinical performance of the new TECNIS Odyssey™ intraocular lens (IOL).

Setting

Fernández-Vega Ophthalmological Institute.

Methods

The clinical study included manifest refraction, uncorrected (UDVA) and corrected (CDVA) distance visual acuities, distance-corrected intermediate (DCIVA) and near visual acuity (DCNVA), binocular photopic and mesopic contrast sensitivity (CS) and binocular defocus curve. Postoperative evaluation was performed 3 months after surgery.

Results

The clinical study involved 122 eyes. The mean postoperative monocular UDVA and CDVA were 0.04 ± 0.05 and 0.01 ± 0.02 logMAR, respectively. All eyes achieved a monocular UDVA of 0.2 logMAR or better and a monocular CDVA of 0.1 logMAR or better. The defocus curve showed a range from +0.50 to -2.5D, in which the mean visual acuity was 0.1 logMAR or better. The binocular distance CS was within normal limits. No intraoperative or postoperative complications related to lens implantation were observed. All patients were satisfied or very satisfied with the procedure.

Conclusions

The implantation of the new TECNIS OdysseyTM IOL, provided a satisfactory range of visual quality from distance to near, in excellent agreement with the optical performance of full depth-of-field intraocular lenses.