ESCRS - POCAT350 - Prospective Three-Month Outcomes Of A Hydrophobic Toric Edof Iol: Visual Performance And Refractive Precision

Prospective Three-Month Outcomes Of A Hydrophobic Toric Edof Iol: Visual Performance And Refractive Precision

Published 2026 - 44th Congress of the ESCRS

Reference: POCAT350 | Type: Free paper | DOI: 10.82333/tc8r-9a53

Authors: Dr Alison Chiu* 1

1University of Sydney ,Sydney,Australia;Pure Sight Eye Surgeons,Sydney,Australia

Purpose

To evaluate visual performance, refractive accuracy, and rotational stability following bilateral implantation of a hydrophobic toric extended depth of focus intraocular lens (ELON IOL 877PETY, Medicontur Medical Engineering Ltd).

Setting

Prospective, single-centre clinical investigation.

Methods

This prospective, single-centre, single-arm interventional study enrolled adults undergoing bilateral cataract surgery with implantation of the ELON 877PETY toric EDOF intraocular lens, an aberration-neutral design extending depth of focus while correcting corneal astigmatism. Assessments at 1, 3, and 12 months included monocular and binocular visual acuity, manifest refraction, defocus curves, contrast sensitivity, and IOL rotational stability under photopic conditions. Patient-reported outcomes were evaluated using Catquest and PRISQ. The current interim analysis focuses on the 3-month outcomes.

Results

At 3 months, binocular DCVA was −0.05 ± 0.06 logMAR, demonstrating excellent distance vision, with functional intermediate vision (UIVA 0.24 ± 0.10; DCIVA 0.23 ± 0.09). Postoperative refraction was highly stable with SEQ tightly centered around emmetropia (97% within ±0.50 D and 100% within ±1.00 D SEQ), with postoperative cylinder clustering close to zero (92% ≤0.50 D; 97% ≤1.00 D). This demonstrates strong refractive predictability.Mean rotation was 2.90° between Day 1 and Month 1, indicating predictable toric alignment. Defocus curves are presented giving further insight into EDOF IOLs. Most patients experienced no or minimal difficulties with their daily activities post-implantation according to the patient reported outcome measures

Conclusions

This prospective clinical evaluation demonstrates that the hydrophobic toric EDOF IOL achieves excellent binocular distance acuity, functional intermediate performance, and high refractive precision with stable rotational alignment. Defocus analysis confirmed an extended functional range of vision, while patient-reported outcomes indicated low levels of visual difficulty. Twelve-month follow-up will to evaluate longer-term stability. This prospective study contributes meaningful clinical data to inform surgical decision-making regarding contemporary EDOF platforms in astigmatic cataract patients.