Haze-Dependent Breakdown Of Pachymetric Agreement Between Scheimpflug Imaging And As-Oct
Published 2026 - 44th Congress of the ESCRS
Reference: FP02.06 | Type: Free paper | DOI: 10.82333/fp79-2e30
Authors: Dr ASIM BURAK GUNDUZ* 1 , Prof. Dr EMINE AKCAY 2
1OPHTHALMOLOGY,ANKARA BILKENT CITY HOSPITAL,ANKARA,Türkiye, 2OPHTHALMOLOGY,ANKARA YILDIRIM BEYAZIT UNIVERSITY,ANKARA,Türkiye;OPHTHALMOLOGY,ANKARA BILKENT CITY HOSPITAL,ANKARA,Türkiye
Purpose
To investigate whether corneal haze severity affects the agreement between Scheimpflug-based pachymetry and anterior segment optical coherence tomography (AS-OCT), and to determine a quantitative haze threshold associated with clinically significant inter-device pachymetric discrepancy.
Setting
Ankara Bilkent City Hospital Department of Ophthalmology, tertiary referral center. Patients with previous corneal cross-linking (CXL) and healthy control eyes were included. Imaging was performed using Scheimpflug tomography (Sirius anterior segment analysis system (CSO, Florence, Italy)) and anterior segment OCT (3D OCT-1 Maestro (Topcon Corporation, Tokyo, Japan)).
Methods
Forty-two CXL eyes and 31 control eyes were evaluated. Thinnest corneal thickness was measured with Scheimpflug imaging and AS-OCT. Inter-device difference (OCT–Sirius) was calculated. Stromal haze was quantified on AS-OCT B-scans using ImageJ by measuring mean pixel intensity in predefined anterior and posterior stromal regions of interest. An anterior/posterior stromal reflectivity ratio (Ant/Post haze index) was computed. Clinical haze grading was recorded. Agreement was assessed with Bland–Altman analysis. Correlation, regression, and ROC analyses were performed. Clinically significant discrepancy was defined as ≥30 µm.
Results
Conclusions
Corneal haze severity-both quantitatively measured by anterior/posterior stromal reflectivity ratio and clinically graded at the slit lamp-significantly reduces agreement between Scheimpflug-derived and AS-OCT pachymetry measurements. Increasing clinical haze grade was strongly associated with widening inter-device discrepancy, indicating a haze-dependent measurement bias rather than random variability. While quantitative reflectivity analysis provides objective confirmation, the mere presence of clinically visible stromal haze should alert clinicians to potential pachymetric inaccuracy with Scheimpflug imaging. In such cases, anterior segment OCT may provide a more reliable assessment of corneal thickness.