ESCRS - PO614 - Trifocal Intraocular Lens With Diffractive-Refractive-Aspheric Technology: Optical Bench Performance And Clinical Results.

Trifocal Intraocular Lens With Diffractive-Refractive-Aspheric Technology: Optical Bench Performance And Clinical Results.

Published 2024 - 42nd Congress of the ESCRS

Reference: PO614 | Type: Free paper | DOI: 10.82333/hmyy-yy32

Authors: Carlos Lisa* 1 , BELEN ALFONSO-BARTOLOZZI 2 , LUIS FERNANDEZ-VEGA CUETO 2 , David Madrid-Costa 3 , JOSE F. ALFONSO 2

1Cornea & Lens,Instituto Universitario Fernández-Vega,Oviedo,Spain;CORNEA AND LENS SURGERY,INSTITUTO UNIVERSITARIO FERNANDEZ-VEGA,OVIEDO,Spain, 2CORNEA AND LENS SURGERY,INSTITUTO UNIVERSITARIO FERNANDEZ-VEGA,OVIEDO,Spain, 3Optica y Optometría,Universidad Complutense de Madridf,Madrid,Spain

Purpose

To assess the performance on the optical bench and the visual results in the clinic of a new trifocal intraocular lens with diffractive-refractive-aspheric technology.

Setting

Instituto Universitario Fernández-Vega (Oviedo), España.

Methods

The Optiflex Trio lens (Biotech, Switzerland) is listed as a trifocal lens with a diffractive-refractive-aspherical design. The study on an optical bench (Polytechnic University of Catalonia) was carried out on a +20.0 diopter (D) lens, according to the ISO (International Organization for Standardization) eye model; Image quality was judged by modulation transfer function (MTF). The clinical study includes a sample of 25 patients (50 eyes) who underwent lensectomy using femtosecond laser-assisted phacoemulsification. The inclusion criteria in the study were: age between 55 and 75 years, and axial length between 22.0 and 26.0 mm. Distance visual acuity and defocus curve were analyzed.

Results

On the optical bench, the lens shows three well-defined foci in the MTF; distance focus, intermediate focus at +1.80 D and a close focus at +3.50 D (3.0 mm aperture diaphragm). In the clinical study, at 12 months, the mean corrected distance visual acuity (LCCA) was 0.94±0.09 (Snellen scale). 87.5% of eyes were within ±0.50 D of spherical equivalent. The mean binocular BCVA at 66, 50, 40 and 33 cm was always greater than 0.9.

Conclusions

On the optical bench, the MTF shows the near focus with an addition of +3.50 D and the intermediate focus of +1.80 D. In the clinic, a defocus curve is obtained with a visual acuity always greater than 0.9, from the distance focus to the near focus (39 cm).