CORRELATION IN ANTERIOR SEGMENT MEASURES BETWEEN A PLACIDO-BASED BIOMETER AND A DUAL PLACIDO AND SCHEIMPFLUG TOMOGRAPHER.
Published 2024 - 28th ESCRS Winter Meeting
Reference: P010 | Type: Presented Poster & Poster | DOI: 10.82333/rcja-7t83
Authors: Alberto Parafita Fernandez* 1 , Julia Barreiro Honrado 2 , Pablo Paz Cotilla 2 , Martin Gonzalez Vazquez 3 , Ariadna Vazquez Prieto 3
1Complexo Hospitalario Universitario de Pontevedra,Pontevedra,Spain;Clinca Oftalmologica Dr Parafita,Ribeira,Spain, 2Complexo Hospitalario de Pontevedra,Pontevedra,Spain, 3Complexo Hospitalario Universitario de Pontevedra,Pontevedra,Spain
Purpose
Setting
Clínica Oftalmológica Dr Parafita, Ribeira.
Methods
Results
Strong correlation was found for K1 (R=0.996), K2(R=0.974), CYL(R=0.969), CCT(R=0.983), ACD(R=0.809) and WTW (R=0.890). Moderate correlation was found for coma (R=0.694), and no correlation was found for Q factor and LSA. Mean values, standard deviations and range for these parameters were Q Sirius -0.33 ± 0.09 (-0.52 to -0.13) vs Q Aladdin -0.14 ± 0.13 (-0.32 to 0.33), LSA Sirius 0.21 ± 0.07 mm (0.06 – 0.42) mm vs LSA Aladdin 0.28 ± 0.13 mm (0 – 0.76) and coma Sirius 0.26 ± 0.13 mm (0.05 – 0.6) vs Coma Aladdin 0.30 ± 0.10 mm (0.06 – 0.49).
Conclusion
There is a strong agreement between both instruments in the main parameters for biometry, pseudophakic lens calculation, laser vision correction assessment and phakic intraocular lens sizing. On the other hand, Aladdin, which is mainly an optical coherence biometer for pseudophakic lens calculation, showed a higher dispersion of values regarding aberrometry and shape factor, which did not correlate well with those obtained by Sirius+.