Comparison Of 2 Scoring Systems For Detection Of Early Forms Of Keratoconus
Published 2023 - 41st Congress of the ESCRS
Reference: FP18.06 | Type: Free paper | DOI: 10.82333/k18h-4r35
Authors: Dishitha Rathod* 1 , Rohit Shetty 2 , Pooja Khamar 2 , Gairik Kundu 2
1Cataracts and Refractive,Narayana Nethralaya,Bangalore,India, 2Refractive,Narayana Nethralaya,Bangalore,India
Purpose
The aim of the study is to investigate the efficacy of the new objective K SCORE method from ANTERION (Heidelberg Engineering GmbH) in comparison to the well established Belin Ambrosio enhanced ectasia display (BAD-D) score from Pentacam AXL Wave (OCULUS Optikgerate GmbH) for the early detection of ectasia susceptible eyes
Setting
This is a prospective study in which 200 eyes of 200 patients undergoing refractive surgery were included (one eye was randomly chosen).
Methods
After taking informed consent, all eyes underwent Pentacam AXL Wave (rotating scheimpfkug technology) and ANTERION (AS-OCT) scanning to derive the respective keratoconus score. The agreement between the parameters used by the two classification softwares were evaluated using interclass correlation coefficient (ICC) and paired simple t-test. Efficacy of the scoring system was assessed by reciever operating characteristics (ROC) curve analysis. All tests were performed using Medcalc 20.115 (MedCalc Software Ltd, Belgium.
Results
Anterior surface keratometry (K) parameters K1, K2 and astigmatism had mean difference and confidence interval of -0.2 [-0.25, -0.16]D, -0.21 [-0.25, -0.16]D and -0.01 [-0.05, 0.03]D, respectively. Similarly, the same for the posterior surface were -0.17 [-0.18, -0.16]D, -0.14 [-0.15, -0.13]D and -0.03 [-0.03, -0.02]D, respectively. The above parameters for both surfaces were significantly different between the devices in the t-test (p<0.001). However, the difference was not clinically relevant (<0.25 D). ICC for the above parameterswas >0.92. ROC analysis of BAD-D had a 94.74% sensitivity and 99.18% specificity. ROC analysis of K SCORE had a 92.86% sensitivity and 91.79% specificity.
Conclusions
Overall the ANTERION and Pentacam AXL Wave parameters showed good agreement. The sensitivity and specificity of the K SCORE were comparable to the BAD-D.