COMPARATIVE ANALYSIS OF SWEPT SOURCE OCT BASED BIOMETER ARGOS AND OPTICAL LOW-COHERENCE REFLECTOMETRY BASED LENSTAR LS 900 FOR TORIC INTRAOCULAR LENS
Published 2023 - 27th ESCRS Winter Meeting
Reference: P092 | Type: Free Paper Submission | DOI: 10.82333/r5kh-z395
Authors: João Romano* 1 , Joana Martins 1 , Nuno Oliveira 1 , João Paulo Sousa 1
1Ophthalmology,Centro Hospitalar de Leiria,LEIRIA,Portugal
Purpose
Setting
Methods
Paired t tests (for normal distribution data) and Wilcoxon tests (for nonparametric data) were employed to compare the difference between the two devices. The Bland-Altman method was implemented to assess their agreement. P values less than .05 were considered statistically significant.
Results
There was no significant difference (p=0.189) between AL measured with the Lenstar (Mean = 23.70 ± 1.69 mm) and Argos (Mean = 23.74 ± 1.22 mm). Both flat keratometry (K1; Mean = 43.16 ± 1.37 D and 43.16 ± 1.46D, respectively) and steep keratometry (K2; Mean = 45.77 ± 1.71 D and 45.69 ± 1.74 D) were also not significantly different (p=0.793 and p=0.275).
Mean ACD was 3.23 ± 0.41 mm on Lenstar and 3.34 ± 0.37 mm on Argos; it was the only measurement statistically significant difference between both biometers (p<0.001). This resulted in a significant difference (p<0.001) in Barrett toric calculation of IOL (Mean = 19.85 ± 3.33 D using Lenstar measures and 20.31 ± 3.30 D using Argos measures).