EVALUATION OF VISUAL QUALITY AND INDEPENDENCE OF OPTICAL CORRECTION, WITH ADVANCED MONOFOCAL TORIC INTRAOCULAR LENSES, RAYONE EMV®TORIC RAYNER. FIRST RESULTS.
Published 2023 - 27th ESCRS Winter Meeting
Reference: P072 | Type: Free Paper Submission | DOI: 10.82333/bsqc-4j29
Authors: Mariano Royo* 1 , Angel Jimenez 1
1Oftalmología,Instituto Oftalmológico de Madrid / Hospital San Rafael,Fuenlabrada,Spain
Purpose
Setting
Visual acuity and blur curve are monitored with the EDTRS optotype at 4 meters and illumination of 85cd / m². Contrast sensitivity using table CSV-1000HGT. At a distance of 2.5 meters (8 feet) with spatial frequencies of 3, 6, 12 and 18 cycles / degree.
Methods
Results
- Monocular UDVA: 0.05 logMAR
- Monocular UIVA: 0.18 logMAR
- Monocular UNVA: 0.48 logMAR
Binocular defocus curve (HOTSPOTS):
- Optical infinity or far vision: 0.05 logMAR
- Intermediate vision 67 cm. (-1.5D): 0.15 logMAR
- Near vision 33 cm. (-3.0D): 0.4 logMAR
Contrast sensitivity test:
- 3 cycles / degree: 5,45
- 6 cycles / degree: 4,52
- 12 cycles / degree: 3,5
- 18 cycles / degree: 2,62
Conclusion
- Monofocal and fully refractive technology.
- Non-pupil dependent functional distance and intermediate visión.
- Positive spherical aberration IOLs maintain stable extended focus.
CLINICAL:
- 100% of patients report independence from glasses in distant and intermediate vision regardless of light conditions.
- 100% of patients can read at 40cm. with an addition of +1.50D to +2.00D.
- In 100% of patients, contrast sensitivity is within normalized values.