EVALUATION OF CORNEAL BIOMECHANICS IN PATIENTS WITH CYSTINOSIS
Published 2023 - 27th ESCRS Winter Meeting
Reference: P011 | Type: Free Paper Submission | DOI: 10.82333/7ahg-vd20
Authors: Puren Isik* 1 , Begum Sulanc 1 , Elif Erdem 1 , Fatma Derya Bulut 2 , Deniz Kor 2 , Cansu Demircan 1 , Inan Harbiyeli 1 , Neslihan Onenli Mungan 2 , Meltem Yagmur 1
1Ophthalmology,Cukurova University,Adana,Türkiye, 2Pediatric metabolism and nutrition,Cukurova University,Adana,Türkiye
Purpose
Setting
Methods
Results
The mean intraocular pressure (IOP) values obtained by Goldmann applanation tonometry, noncontact tonometry, and Corvis ST® in patients with cystinosis were 14±1.4, 16.6±1.8, and 16.2±2 mmHg, respectively. IOP values obtained by three different methods were found to be similar (p=0.152). First applanation velocity (A1V), deformation amplitude (DA), DA ratio (DAR) and integrated radius (IR) were found to be statistically significantly higher in patients with cystinosis compared to the control group (p=0.001, p<0.001, p=0.004, p<0.001, respectively). On the other hand, the second applanation velocity (A2V), first applanation stiffness parameter (SP-A1), central corneal thickness (CCT), stress-strain index (SSI), and peak distance (PD) at the highest concavity point were found to be low in patients with cystinosis (p=0.001, p<0.001, p=0.003, p=0.002, p=0.003, respectively).
While there was a negative correlation between age and the PD at the highest concavity point (r= -0.336, p=0.034), no significant correlation was found with other parameters (p>0.05 for all).