SCLERAL RIGIDITY INFLUENCES CORNEAL BIOMECHANICS
Published 2023 - 27th ESCRS Winter Meeting
Reference: P006 | Type: Presented Poster & Poster Submission | DOI: 10.82333/gkgt-8780
Authors: Alejandra Consejo* 1 , Ana Ramos-Arizcuren 1 , Elke O. Kreps 2 , Amaya Perez del Palomar 3
1Applied Physics,University of Zaragoza,Zaragoza,Spain, 2Ophthalmology,Ghent University Hospital ,Ghent,Belgium, 3Mechanical Engineering,University of Zaragoza,Zaragoza,Spain
Purpose
Setting
Methods
A three-dimensional finite element model including the cornea, the anterior sclera and the vitreous was implemented. These materials are defined as hyperelastic, following the parameters accessible from the literature, and applying a Holzapfel and Yeoh model for the cornea and the sclera, respectively. To achieve a realistic model, corneal fibers were taken into consideration. To mimic the mechanical stimulation of the cornea, an air-puff was simulated following the characteristics of commercially available tonometers. Intraocular pressure was fixed to 18 mmHg. Once the model was completed, scleral rigidity was modified from 0.4 MPa (soft sclera) to 1.6 MPa (hard sclera), and the corresponding maximum corneal deformation amplitude was evaluated.
Results
Conclusion