VISUALIZATION OF LIGHT DISTRIBUTION AND RAY PROPAGATION IMAGING OF EXTENDED-DEPTH-OF-FOCUS IOLS
Published 2023 - 27th ESCRS Winter Meeting
Reference: O32 | Type: Free Paper Submission | DOI: 10.82333/zd87-4b95
Authors: Ramin Khoramnia* 1 , Isabella Baur 1 , Gerd Auffarth 1 , Grzegorz Łabuz 1
1University Eye Clinic Heidelberg,Heidelberg,Germany
Purpose
Setting
Methods
We compared the non-diffractive AcrySof IQ Vivity (Alcon Inc.) to the diffractive Symfony ZXR00 (Johnson & Johnson Vision) and AT Lara 829 MP (Carl Zeiss Meditec) in a laboratory investigation. The monofocal AcrySof IQ SN60WF (Alcon Inc.) served as a control. The point spread function (PSF) as a measure of spurious light was applied to assess the risk of photic phenomena. Each lens was submerged in a fluorescein solution and a 520-nm laser light was projected through the IOL. The emerging light was captured using a camera mounted on a microscope.
Results
Evaluation of the PSF showed the lowest level of light spread for the monofocal IOL. The Acrysof IQ Vivity showed minimally extended light spread and both diffractive lenses showed excessive light intensity around the center of the PSF with several halo-like rings. The ray visualization and light intensity profile of the monofocal IOL revealed one distinct focus corresponding to the IOL nominal power, while the Vivity showed an elongated focus. For both diffractive EDoF IOL models we observed two foci for far and intermediate distance, with the Symfony demonstrating a slightly higher peak of the light intensity profile at the intermediate focus and the AT Lara showing far-dominance.
Conclusion
Ray propagation imaging confirmed an extended range of focus for all EDoF IOL models. Only the non-diffractive EDoF IOL showed a continuous elongated focus. The Vivity was associated with less spurious light in the PSF. This might translate to a lower level of photic phenomena in clinical use than with the diffractive lens designs.