Refractive Cataract Surgery With A New Toric Intraocular Lens With An Enhanced Optical Profile And Modified Haptics
Published 2022
- 40th Congress of the ESCRS
Reference: PP22.14
| Type: Free paper
| DOI:
10.82333/s3ba-9e08
Authors:
Oege Goslings* 1
, Nic Reus 2
1Ophthalmology,Elisabeth TweeSteden Hospital,Tilburg,Netherlands, 2Ophthalmology,Amphia Hospital,Breda,Netherlands
Purpose
To present clinical data on residual refractive astigmatism, visual acuity (UDVA, CDVA) and toric alignment of the Tecnis Eyhance Toric II intraocular lens IOL (J&J Vision) after implantation in cataract surgery
Setting
Departments of Ophthalmology, St. Elisabeth TweeSteden Hospital, Tilburg and Amphia Hospital, Breda, The Netherlands
Methods
83 eyes were implanted with a Tecnis Eyhance Toric II IOL with different cylinder powers at the IOL plane (DIU100 n=10, DIU150 n=22, DIU225 n=14, DIU300 n=7, DIU375 n=16, DIU450 n= 16, DIU525 n=6). Outcome measures included residual refractive astigmatism, (un-)corrected distance visual acuity and toric IOL alignment
Results
At 30-90 days postoperatively, the residual refractive cylinder was -0.30 ± 0.33 (DIU100), -0.33 ± 0.30 (DIU150), -0.29 ± 0.26 (DIU225), -0.58 ± 0.29 (DIU300), -0.55 ± 0.46 (DIU375), -0.66 ± 0.26 (DIU450), -0.46 ± 0.33 (DIU525) and -0.42 ± 0.39 (DIU groups combined). UDVA was 0.10 ± 0.13 (logmar) and DCVA was 0.01 ± 0.08 for the DIU groups combined. Degree misalignment was 0.8 ± 1.8 (DIU100), 0.6 ± 1.2 (DIU150), 1.1 ± 2.2 (DIU225), 0.6 ± 0.8 (DIU300), 1.8 ± 1.83 (DIU375), 1.63 ± 2.13 (DIU450), 2.8 ± 4.3 (DIU525) and 1.07 ± 1.83 (range: 0 – 9) for the DIU groups combined
Conclusions
Implantation of the Tecnis Eyhance Toric II IOL seems to be an effective method for astigmatism correction in cataract patients. The residual refractive cylinder remains low even with increasing cylinder powers up to 5 diopter because of an accurate toric alignment profile of this new lens with enhanced optical profile and modified haptics