ESCRS - PO429 - Comparative Analysis Of 10-2 Elisar Standard Algorithm Of Advanced Vision Analyzer With Sita Standard Of Humphrey Field Analyzer

Comparative Analysis Of 10-2 Elisar Standard Algorithm Of Advanced Vision Analyzer With Sita Standard Of Humphrey Field Analyzer

Published 2022 - 40th Congress of the ESCRS

Reference: PO429 | Type: Free paper | DOI: 10.82333/a4ca-t498

Authors: Priya Narang* 1 , Amar Agarwal 2

1Cataract and Refractive surgery,NARANG EYE CARE & LASER CENTRE,Ahmedabad,India, 2Cataract and Refractive surgery,Dr. Agarwal's Eye Hospital,Chennai,India

Purpose

To prove equivalence of Advanced vision analyser’s 10-2 Elisar standard algorithm (ESA) with 10-2 SITA Standard (SS) of Humphrey Field Analyzer (HFA).

Setting

Narang Eye Care & Laser Centre, India; Dr. Agarwal's Eye Research Centre, India

Methods

All subjects enrolled in the study underwent a 10-2 test with AVA and HFA. After implementing the exclusion and inclusion criteria, the threshold estimates of 112 eyes of 107 subjects [64 Glaucoma subjects (66 eyes);35 healthy controls (36 eyes); 9 glaucoma suspect subjects (10 eyes)] were analysed. Pointwise sensitivity analysis, Intra class correlation (ICC) and comparison with Bland Altman plots were done to compare the 10-2 threshold estimates obtained with both devices.  

Results

Bland Altman plot showed significant correlation for mean sensitivity (MS), mean deviation (MD) and pastern standard deviation (PSD) values for both devices. Correlation coefficient for pointwise threshold values strongly correlated for both the devices.

For MS, overall ICC value was 0.96 (P<0.001) with MB of 0.0 dB and limits of agreement (LOA) range of 7.59.

Conclusions

Statistical results denote adequate equivalence between AVA and HFA as threshold estimates of AVA strongly correlate with HFA for 10-2 program.