First Real World Experience With A New Multifocal Iol With Extended Depth Of Focus And Violet Light Blocker
Published 2022 - 40th Congress of the ESCRS
Reference: PO226 | Type: Free paper | DOI: 10.82333/6xh7-dm28
Authors: Oege Goslings* 1
1Ophthalmology,Elisabeth TweeSteden Hospital,Tilburg,Netherlands
Purpose
To present clinical data on visual acuity, defocuscurve and photopsia profile of the new Tecnis Symfony Plus Optiblue intra ocular IOL (J&J Vision)
Setting
Department of Ophthalmology, Elisabeth TweeSteden Ziekenhuis, Tilburg, The Netherlands
Methods
10 eyes have been implanted with a Tecnis Symfony Plus Optiblue intra ocular IOL. Outcomes measures include visual acuity (UDVA, CDVA, UIVA, DCIVA, UNVA, DCNVA), defocus curve and photopsia profile
Results
(preliminary) a broad defocus curve is shown : Snellen Visual Acuity / Diopter defocus is : +1/0.38, +0.5/0.68, plano/0.98, -0.5/0.98, -1/0.98, -1.5/1.0, -2/0.98, -2.5/0.83, -3/0.70. UDVA, DCVA: 1.0+, UIVA, DCIVA: 1.0, UNVA, DCNVA, Jaeger 2. Photopsia profile: slight starburst phenomena
Conclusions
The new Tecnis Symfony Plus optiblue IOL shows improved visual acuity values and enhanced defocus profile compared to its predecessor Tecnis Symfony IOL. Photopsia were perceived as minor disturbant phenomena