Comparison Of Anterior Segment Measurements With Multiple Devices And A New Multifunctional Unit For Preoperative Examination Of Cataract Surgery
Published 2022 - 40th Congress of the ESCRS
Reference: PO130 | Type: ESCRS 2022 - Posters | DOI: 10.82333/6wpb-3009
Authors: Bo Yi Kim* 1 , Ikhyun Jun 1
1Ophthalmology,Yonsei University College of Medicine,Seoul,Korea, Republic Of
Purpose
Setting
Methods
With above six different devices, the mean keratometry (Km), central corneal thickness (CCT), white-to-white (WTW) distance, and intraocular pressure (IOP) values were measured and compared. In comparing keratometry, MR-6000, KR-1, Pentacam HR, IOLMaster 700, and Orbscan II were used. In evaluating CCT, MR-6000, Pentacam HR, and Orbscan II were included. WTW was measured by MR-6000, IOLMaster 700, and Orbscan II. The value of IOP was compared by MR-6000 and FT-1000. As statistical analysis for evaluating correlation and agreements between devices, repeated measures analysis of variance, Wilcoxon signed rank test, intraclass correlation coefficient (ICC), and Bland–Altman plot were used.
Results
Thirty eyes of 30 patients were evaluated. Statistically significant differences in Km between MR-6000, KR-1, Pentacam HR, and IOLMaster 700 were not observed (p > 0.05). Including Orbscan II, all five devices had almost perfect agreement in measuring keratometry (ICC > 0.80, p < 0.05). CCT measured by MR-6000 was significantly different from that of Pentacam HR and Orbscan II measurements (p < 0.05). Substantial correlation with that of Pentacam HR and Orbscan II measurements were observed(ICC > 0.60, p < 0.05). The WTW distance measured by MR-6000 was not significantly different from that measured by IOLMaster 700 with agreement but was different from that measured by Orbscan II. IOP measured by MR-6000 was not correlated with FT-1000.
Conclusions
MR-6000 has strengths in measuring different parameters at once on the spot. Therefore, we expected it to unify or substitute multiple preoperative examinations for cataract surgery. As a result, keratometric values obtained through MR-6000 can be interchangeably used with other devices based on good correlation and agreement. However, the CCT, WTW, and IOP values were not interchangeable with a single multifunctional unit for cataract surgery preoperative examination.