ROTATIONAL STABILITY AND CENTRATION OF A TORIC LENS USING THE ITRACE TECHNOLOGY
Published 2022 - 26th ESCRS Winter Meeting
Reference: P080 | Type: Abstracts | DOI: 10.82333/g36q-qc23
Authors: Jesús María Carpintero Sánchez* 1 , Ana Hervas Ontiveros 1 , Enrique Garcia Soler 1 , Sara Fathi Nieto 1 , Vanesa Osorio Alayo 1 , Amparo Ortiz Seller 1 , Andrea Llovet Rausell 1
1Oftalmología,Hospital La Fe,Valencia,Spain
Purpose
intraocular lens (IOL) (AST, Boston, Masschusets) using objective image and
Itrace (Tracey, Houston, Texas) technology.
Setting
Methods
astigmatism in biometrics > 3 Prismatic Diopters (PD), presenting cataract
surgery were implanted monocularly with the Asqelio Toric Monofocal IOL and
followed up to 90 days postoperatively. Images of the IOL were captured with
the Itrace technology through a maximally dilated pupil immediately after
surgery and after 1 to 3 days, 30 days, and 90 days. Centering of the IOL
were analyzed with Itrace technology through angle kappa and alpha
distance and the tilt of the IOL through the measurement of the coma.
Results
kappa positive in all cases, ranging from +1° to +5°. The mean angle kappa
was 2.68° and 2.10° in the right and left eye, respectively. Different subjective
perception of photic phenomena between the two eyes was recorded only in
2 patients. Wave aberration measurements of the same IOL were highly
repetitive (deviation <0.02 μm). Coma increased with decentration at a rate of
0.18 μm / mm and 0.19 μm in the right and left eye, respectively.
Conclusion
implantation. The measurements with the Itrace technology suppose a perfect
system for evaluate the postoperative centering and the tilt of the lens
assuming an improvement of the visual quality of the patient.