EVALUATION OF SELECTED BIOMETRIC PARAMETERS IN CATARACT PATIENTS – A COMPARISON BETWEEN TWO SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY-BASED BIOMETERS
Published 2022
- 26th ESCRS Winter Meeting
Reference: FP43
| Type: Abstracts
| DOI:
10.82333/31qd-g026
Authors:
Ewa Mrukwa-Kominek* 1
, Mateusz Porwolik 1
, Monika Sarnat-Kucharczyk 1
1Department of Ophthalmology, Medical University of Silesia,Department of Ophthalmology, Medical University of Silesia,Katowice,Poland
Purpose
To compare biometry of eyes obtained with two swept-source optical coherence tomography-based biometers: Argos (A) (Movu Inc., Komaki, Japan) and IOLMaster 700 (IM) (Carl Zeiss Meditec, Jena, Germany).
Setting
Department of Ophthalmology, Faculty of Medical Sciences in Katowice, Medical University of Silesia, Katowice, Poland
Department of Ophthalmology, Professor K. Gibinski University Clinical Center, Medical University of Silesia. Katowice, Polandk
Methods
Biometry of 105 eyes of 105 patients before cataract surgery with implantation of intraocular lens were analyzed in this study. Following parameters such as axial length (AL), anterior chamber depth (ACD), lens thickness (LT) from both devices were compared. According to the axial length measurements patients were divided into three groups: group 1 – short eyes (AL <22.5mm), group 2 – average eyes (AL ≥ 26.0 mm) and group 3 – long eyes (22.5 ≤ AL ≤ 26.0 mm). Data were analyzed using Statistica 13.3 software, StatSoft
Results
Statistical significance in axial length was indicated in group of short eyes (n = 26) - mean AL (A) 21,90mm (± 0,59mm) vs AL (IM) 21,8mm ± (0,61mm) (p< 0,001) and in group of long eyes (n = 5) - mean AL (A) 27,95mm (± 2,62mm) vs mean AL (IM) 28,10mm (± 2,64) (p<0,05).
In group of average eyes (n = 74) outcomes were similar - mean AL (A) 23,56mm (± 0,70mm) vs mean AL (IM) 23,56mm (± 0,71mm) (p> 0,05). Anterior chamber depth measurements were higher obtained with Argos than with IOLMaster 700 - mean ACD (A) 3,06mm (±0,48mm) vs mean ACD (IM) 2,92mm (±0,46) p< 0,001
There was no statistical significance in mean LT: mean LT (A) 4,75mm (±0,46mm) vs mean LT (IM) 4,72mm (±0,44mm) (p= 0,054).
Biometry of one eye with dense cataract could be measured only with Argos using Enhanced Retinal Visualization mode
Conclusion
Axial length measurements with both devices were different in groups of short and long eyes, but comparable in group of average eyes. Anterior chamber depth values obtained with Argos were higher than measurements acquired with IOLMaster 700. Lens thickness measurements were similar with both devices. Eyes with dense cataract could be measured with higher accuracy with Argos than with IOL-Master 700