Comparison Of Two Specular Microscopes Devices In Endothelial Clinical Parameters
Published 2023
- 41st Congress of the ESCRS
Reference: PO0247
| DOI:
10.82333/0634-bt19
Authors:
Ruiling Xi 1
, Yinglin Zhang 2
, Lingxi Zeng 3
, Lingxi Hu 4
, Risa Higashita 1
, Zheng Chang 5
, Chen Lin 6
, Jiang Liu* 1
1Research Institute of Trustworthy Autonomous Systems and Department of Computer Science and Engineering,Southern University of Science and Technology,Shenzhen,China, 2School of Computer Science, University of Nottingham Ningbo China,Ningbo,China, 3Computer Science and Engineering,Southern University of Science and Technology,Shenzhen,China, 4School of Computer Science,University of birmingham,Birmingham,United Kingdom, 5Department of Cornea,Shenzhen Aier Eye Hospital Affiliated to Jinan University,Shenzhen,China, 6ophthalmology department,Shenzhen People's Hospital,Shenzhen,China
To evaluate the intra-device repeatability and inter-device agreement of endothelial clinical parameters obtained with two specular microscopes.
Shenzhen Aier Eye Hospital, Shenzhen, China.
110 eyes of 55 healthy subjects were included in this cross-sectional study. Average age was 29.9 ± 5.86, and 33 subjects were female. Each eye was captured twice by EM-4000 (Tomey, Japan) and SP-1(Topcon, Japan) on the same visit separately. Endothelial Cell Density (ECD) was recorded by their built-in analysis software. To avoid software-dependent bias, we also introduced a deep learning network with our developed parameter analyzer on captured specular microscope images. Intra-device repeatability and inter-devcie agreements were evaluated using Bland-Altman plots and intraclass correlation (ICC), respectively.
The ECD of EM4000 and SP-1 by the built-in analysis are 2786.89 ± 230.68 and 2977.90 ± 336.87 respectively, while by ourself-developed analyzer are 2996.06 ± 276.72 and 3370.90 ± 317.58. The Bland-Altman plot method demonstrates that the zero difference fell within 95% limits of agreement (LoA) for all experiments. EM-4000 has narrower LoA compared with SP-1, considering both captured image and build-in software analysis. The inter-device measurement agreement between two devices is poor on captured image, with ICC = 0.02, P = 0.89, and r = 0.01. The inter-device measurement agreement between two devices is moderate through build-in analysis software, where ICC = 0.54, P < 0.01, and r = 0.39.
Continuous and proper monitoring of ECD is crucial after cataract refractive surgery to ensure surgical outcomes and detect potential complications. Therefore, a reliable and consistent method for measuring ECD is critical. In this study, we compared two specular microscope devices for ECD estimation. Both EM-4000 and SP-1 showed intra-device repeatability. EM-4000 demonstrated consistently smaller variance in both captured image and built-in software analysis, which might suggest higher stability. Other factors, such as SP-1's higher ECD estimation mean value, may also contribute to its higher variance. The agreement between the two instruments is moderate. Therefore, we do not recommend interchangeable use of EM-4000 and SP-1 microscopes.