Prospective, Randomized, Multicenter Clinical Evaluation Comparing A Monofocal Enhanced Depth Of Focus Intra Ocular Lens With A Monofocal Intra Ocular Lens
Published 2023 - 41st Congress of the ESCRS
Reference: FP17.09 | Type: Free paper | DOI: 10.82333/2j6h-v778
Authors: Mark Packer* 1
1Packer Research Associates,Boulder,United States
Purpose
To investigate the safety and effectiveness of a monofocal enhanced depth of focus intraocular lens (xact Mono-EDoF IOL ME4, Santen Pharmaceuticals) when compared to a standard monofocal intraocular lens (Tecnis Monofocal ZCB00, Johnson & Johnson Vision).
Setting
Multicenter clinical investigation conducted at 3 sites (Servico de Oftalmologia, Hospital de Braga, Braga, Portugal; Department of Ophthalmology, University Hospital Pellegrin, Bordeaux, France; Oxford Eye Hospital, Oxford, UK).
Methods
Results
124 eyes of 62 patients (45 female, 17 male) were implanted. Mean age was 69.9 ± 7.0 years (52 - 84). Analysis includes 32 eyes at 3 months and 20 eyes at 6 months in the Tecnis ZCB00 group and 62 eyes at 3 months and 46 eyes at 6 months in the Mono-EDOF ME4 group. Mean MRSE and cylinder at 6 months measured 0.03 ± 0.59 D and -0.26 ± 0.79D in the Tecnis group and -0.14 ± 0.32 D and-0.25D ± 0.59D in the ME4 group. Mean monocular and binocular UCDVA measured logMAR 0.09 ± 0.14 and -0.02 ± 0.11 in the Tecnis group and 0.02 ± 0.13 and -0.04 ± 0.10 in the ME4 group. UCIVA and DCIVA measurements indicated approximately 1 line better acuity in the ME4 group. Monocular and binocular defocus curve testing indicated 1.0 and 1.5 D depth of focus.
Conclusions