Dermot McGrath
Published: Monday, May 13, 2019
[caption id="attachment_15046" align="alignleft" width="640"]

Gilles Lesieur MD[/caption]
Intraocular lens design was found to be the key determining factor in the rate of posterior capsular opacification (PCO) development in a large series of microincision cataract surgery (MICS) lenses, according to a study.
At the meeting of the French Society of Ophthalmology, Gilles Lesieur MD presented the analysis of a large series of 6,500 lenses with up to eight years’ follow-up.
“Although we know that PCO is multifactorial and that surgical technique, IOL biomaterial and patient characteristics may all play a role, the key factors in its development seem to be the presence of a square-edged design and the sharpness of the posterior edge of the optic,” he said.
Dr Lesieur’s retrospective analysis included seven different MICS lens types: CT Asphina (Zeiss), Micro AY, MicroSlim and Micropure (PhysIOL), Akreos MICS (Bausch and Lomb), Incise (Bausch and Lomb) and Synthesis (Cutting Edge).
The Akreos MICS was found to have the highest rate of Nd: YAG laser capsulotomy over the eight-year follow-up period. “This is perhaps due to the design of the lens, with excessive polishing of the optic edges presenting less of a barrier to lens epithelial cell migration,” he said.
The MicroSlim, with its uninterrupted square-edge optic, recorded the lowest rate of PCO, noted Dr Lesieur, while the CT Asphina’s plate haptic design with a square edge only delayed PCO for the first three years with Nd:Yag rates increasing thereafter.
Tags: microincision cataract surgery, posterior capsular opacification
Latest Articles
Addressing Postoperative Visual Complications
Managing aberrations after laser refractive surgery requires a multi-layered approach.
Read more...
3D Printing Helps Transform Ukrainian Eye Care
The country’s ophthalmologists offer valuable experience in treating ocular trauma and prosthesis design.
Read more...
Winning Essay Says ‘Collective Desire’ Must Drive DEI Implementation
Read more...
Emerging Microbial Trends That Could Affect Your Practices
A triptych of challenges paints a concerning picture for ophthalmologists across the globe.
Read more...
Improving Outcomes with Laser-Assisted Surgery
Femtosecond laser offers a multifunctional tool for improving the safety and efficacy of cataract and refractive lens exchange procedures.
Read more...
Could the Corneal Transplant Pool Increase?
Modifying or discarding major contraindications for keratoplasty could mean more patients have their sight restored.
Read more...
Matching Premium IOLs to Visual Lifestyles
From monofocal to full-range solutions, each practice needs comprehensive understanding.
Read more...
Going Dutch on Acanthamoeba Keratitis
A world-first trial suggests a new medication could beat the disease.
Read more...
Avoiding Intracorneal Ring Segment Complications
Femtosecond lasers are helping improve refractive results with fewer problems.
Read more...
Fine-Tuning IOLs After Surgery
Lens power adjustment technology is developing quickly.
Read more...