Amsterdam 2013 Programme Satellite Meetings Registration Exhibition Virtual Exhibition Hotels Visa Letter Invitation
Search Abstracts by author or title
(results will display both Free Papers & Poster)

Moria One Use-Plus SBK (sub-Bowmans keratomileusis) head: a useful tool in the refractive surgeon’s armamentarium

Poster Details

First Author: S.Abd El Wahab KUWAIT

Co Author(s):    M. El Fayoumi              

Abstract Details



Purpose:

To study the safety efficacy and advantages of the Moria One Use-Plus SBK head.

Setting:

Al Salam international Hospital, Kuwait.

Methods:

500 eyes of 250 patients underwent LASIK surgery using the One Use-Plus SBK head for creation of an ultrathin LASIK flap. In follow up at 1, 6, 12 and 24 months, both uncorrected visual acuity (UCVA) and best corrected visual acuity (BCVA) were recorded. A keratoconus screening corneal topography was done on the last follow up visit.

Results:

500 eyes of 250 myopic patients were included in this study, 115 males, 135 females, mean age was 27.5 years range (18 – 44). The mean preoperative refractive error was -5.0 diopters sphere, range (-3 to -9), mean cylindrical error was -2.5 diopters cylinder, range (-0.75 to -4.5). The mean UCVA was 0.07 range (0.01 -0.3), it improved to 1.0, range (0.8-1.25) on day one post operatively. At the end of the follow up, the mean UCVA was 1.0 and the mean BCVA was 1.0. The mean stromal residual bed thickness was 362ṁ, range (304ṁ to 466ṁ) the mean central flap thickness was 102ṁ, range (82 - 120). Complications were perilimbal bleeding in 15 eyes, fine irregular stromal edge cuts at the sides of the hinge in 10 eyes, no epithelial defects, no epithelial heaping or sliding, no irregular stromal bed surface cuts, no free caps or button holes or incomplete flaps, DLK occurred in one eye of one patient, no slipped flaps, macro or microstriae, or epithelial down growth were seen over the follow up period. At the end of the follow up period none of our patients had topographic evidence of keratoconus.

Conclusions:

The One Use-Plus SBK head, is a safe and effective method for creation of an ultrathin LASIK flap

Financial Disclosure:

NONE