Optical quality of the Visian implantable collamer lens for different refractive powers

Poster Details

First Author: R.Montés-Micó SPAIN

Co Author(s):    C. Pérez-Vives   T. Ferrer-Blasco   A. Domínguez-Vicent   D. Monsálvez-Romín     

Abstract Details



Purpose:

Optical Quality of the Visian Implantable Collamer Lens for Different Refractive Powers

Setting:

University of Valencia

Methods:

Wavefront aberrations of the -3, -6, -9, -12 and -15 diopters (D) V4b ICLs were measured at 3- and 4.5-mm pupils. The root mean square (RMS) of total higher order aberrations (HOAs), trefoil, coma, tetrafoil, secondary astigmatism, and spherical aberration were evaluated. In addition, Modulation Transfer Function (MTF) of the five ICL powers was measured for a 3-mm pupil. The Point Spread Functions (PSFs) of each ICL evaluated was calculated from the wavefront aberrations at 4.5-mm pupil.

Results:

The ICLs evaluated had negative spherical aberration and negligible amounts of other aberrations. The negative spherical aberration increases when the ICL power increases being related with its innate optical properties. At 3-mm pupil no statistically significant differences between ICLs were found for all the Zernike coefficient RMS values analyzed (p>0.05). At 4.5-mm pupil, significant RMS values for the spherical aberration and total HOAs were found between medium-low and high powers (p<0.05). Similar MTFs were obtained for all ICLs, although they slightly worsened when increased the ICL power.

Conclusions:

ICLs evaluated provide good optical quality in terms of wavefront aberrations, MTF and PSF. Although, spherical aberration increases with ICL power these values are clinically negligible to affect the visual quality after its implantation.