Amsterdam 2013 Programme Satellite Meetings Registration Exhibition Virtual Exhibition Hotels Visa Letter Invitation
Search Abstracts by author or title
(results will display both Free Papers & Poster)

A study of the repeatability of GCC thickness (ganglion cell complex) measurements using Optovue RT-100 V 6.3 OCT machine

Poster Details

First Author: P.Jegadeesan INDIA

Co Author(s):    S. Ratnasababathy   G. Parasuram           

Abstract Details



Purpose:

It is a known fact that ganglion cell loss precedes RNFL loss in patients with early Glaucoma. The purpose of our study is to know the repeatability of the ganglion cell thickness measurements with the OPTOVUE RT-100 OCT machine.

Setting:

Sundar Eye Hospital, India

Methods:

10 consecutive readings of Ganglion cell thickness measurements were done for 6 eyes of 3 people . The program included GCC average, Superior, Inferior, Superior-Inferior, FLV .

Results:

The standard Deviation ranged from 0.62 and 1.44microns.

Conclusions:

The repeatability of the GCC thickness with the Optovue RT-100 machine appears to be good with low SD.

Financial Disclosure:

NONE

Back to previous