Amsterdam 2013 Programme Satellite Meetings Registration Exhibition Virtual Exhibition Hotels Visa Letter Invitation
Search Abstracts by author or title
(results will display both Free Papers & Poster)

Bladeless femto-cataract procedure with the new high frequency FEMTO LDV Z6 PowerPlus: first clinical outcomes

Poster Details

First Author: B.Pajic SWITZERLAND

Co Author(s):                  

Abstract Details



Purpose:

To assess higher predictability of the customize cut precision and dimension with a smooth cut and higher impact of clinical outcome

Setting:

Swiss Eye Research Foundation, Eye Clinic Orasis, Reinach AG, Switzerland

Methods:

Performance of the FEMTO LDV Z6 PowerPlus is evaluated in vivo for cataract surgery. 16 eyes were analyzed following cataract surgery. Analysis was done using topography, wavefront measurement, corneal OCT, confocal corneal laser-scanning microscopy and clinical examinations.

Results:

Achieved very smooth cutting surface by application of small bubbles and high frequency. A perfect access was achieved of main, primary incision and paracentesis. The main incision access diameter shape of 2.2ḟ0.05mm and paracentesis of 0.8ḟ0.02mm were seen. The incisions were done topography matched with a precision of the target angle of ḟ2.5Ḟ.

Conclusions:

FEMTO LDV Z6 PowerPlus, which combines pulse energy in low nanojoule and frequency in megahertz range, results of a highly predictable quality resection for cataract surgery.

Financial Disclosure:

NONE

Back to previous